PLEASE BE AWARE: This device is only available on loan!

The Bio-Impulser PMS10 is a therapeutic device for the treatment of neurological clinical pictures.

The therapeutic device allows treatment using time-controlled EMG biofeedback training, electrostimulation, a combination of both therapeutic methods and EMG-triggered electrostimulation. In addition to the therapeutic functions, the device also has intuitive operation and a high level of user-friendliness.

The different treatment programs support the active training of endurance, strength and co-ordination of the musculature to recover patterns of movement (such as in hemiparesis after a stroke) and in the treatment of spasticity.

The electrostimulation programs provide additional strengthening of the musculature.
The muscle activity is recorded using adhesive electrodes and displayed on an LCD screen.

The course of therapy can be documented and checked by saving the exercise data. Ideally, exercise evaluation is carried out using computer software, which also allows treatment parameters to be adjusted if necessary.

  

  • Areas of application
  • Functions
  • Package
  • Optional accessories
  • Dimensions and weight
  • Display
  • Memory
  • Power supply
  • EMG settings & voltage measuring range
  • Setting times for exercises

Spasticity or incomplete paralysis after a stroke or skull/brain trauma
Lesions in the central nervous system (operative interventions, inflammations, traumatic injuries)
Muscular hypotrophy and atrophy
Paralysis (as a result of diseases to peripheral nerves capable of regeneration)
Muscular development & rehabilitation

EMG biofeedback training
Electrostimulation
Combined therapy (EMG / electrostimulation)
EMG-triggered electrostimulation

1x "PMS10" therapeutic device
1x sturdy plastic case
1x charger
1x operating instructions
Further optional accessories are required, dependent on the treatment

Dimensions 190x100x60mm
Weight 525g

LCD screen

Data memory for max. 75 exercises or 8.5 hours

NiMH rechargeable battery, 4x R6 round cells with at least 700mAh / 1.2V per battery
Chager for charging the batteries

EMG voltage range 0 - 300µV
Resolution of 0.04µV to 3µV (depending on EMG range selected)
EMG range is set manually

Tension time 0 - 200s
Relaxation time 1 - 200s
Stimulation time 1 - 200s
Pause 0 - 200s



Contraindications

Simultaneous connection to a surgical device with a high frequency can lead to burning of the skin at the electrode contact points. The size of the electrodes used must be tailored according to the indication. Do not operate the device in the presence of flammable gases. Only use the device if safe handling is guaranteed and dangers can be ruled out. The commissioning of the device close to therapeutic short-wave or microwave devices can lead to instability in the initial values. Never apply the electrodes to tumours or broken skin. In the case of irritation to the skin or mucous membranes, stop treatment and inform the doctor in attendance immediately.

Classification under the Medical Devices Act

Class IIa medical device